Sterilization - Dentsply Sirona Midwest T1 Energo Instructions For Use Manual

Table of Contents

Advertisement

Available languages

Available languages

7 Reprocessing

7.3 Sterilization

18
7.3
Sterilization
CAUTION!  Risk of contamination! Make sure you observe the
following intervals.
Intervals
● Prior to initial operation
● Prior to each subsequent use
CAUTION!  Risk of contamination! Steam sterilizers are approved
that comply with EN 13060, class B or EN 13060, class S, and which
are also suitable for sterilizing instruments.
Wrapped sterilization
ü The instrument is clean  [→ 15].
ü The instrument is lubricated  [→ 16].
ü The instrument is packed in packaging suitable for sterilization and
storage (FDA cleared, paper/plastic composite packaging according
to ISO 11607).
➢ Sterilize the instrument in the steam sterilizer with saturated water
vapor.
Temperature:
Overpressure:
Packaging
Steam sterilization
Wrapped
Gravitation
process
Wrapped
Gravitation
process
Wrapped
Dynamic Air Re-
moval
Wrapped
Dynamic Air Re-
moval
NOTICE!  Do not exceed 140 °C (284 °F) even during the drying phase.
After sterilizing
1.
CAUTION!  The instrument may still be hot.
NOTICE!  Do not attempt to accelerate the cooling process by
immersing the instrument in cold water.
2. Store all instruments so that they are protected from contamination.
3. Sterilize again once the storage period has elapsed.
Instructions for use Midwest
132 °C (270 °F) / 135 °C (275 °F)
1.87 bar (27.12 psi) / 2.13 bar (30.89 psi)
Holding time
at least
15 minutes at 132 °C
10 minutes at 135 °C
4 minutes at 132 °C
3 minutes at 135 °C
D3751.201.03.02.02    2023-11
Dentsply Sirona
®
T1/T2 Energo™
Drying time
at least
15 minutes
30 minutes
20 minutes
16 minutes
68 19 200 D3751