5.2.4
Input operation port setting
5
Set the input operation port to the serial port when operating the STP pump via the X3 COM1 or X5
STP-LINK.
Set the parameter of the "Input operation port" to the serial port which operates STP pump, following
Section 5.4.16. The parameter value of the factory setting is "I/O REMOTE" (parallel port). The "Input
Operation Port" can also be changed via the STP-Link (optional accessory) or the display unit iDT-
001 or iDT-002 (optional accessory).
Parallel port
Serial port
Note:
Any commands other than STP pump operation are effective in every port regardless of the
input operation port setting.
MT-94E-001-C
Page 76
STP-iS1607 Series Turbomolecular Pump
Input operation
port
I/O Remote
COM1
X3 COM1 connector, Profibus (optional accessory)
STP-LINK
Table 18 - Input operation port
Remark
X2 REMOTE connector
X5 STP-LINK connector