Readcounters - Edwards STP-iS1607 Series Instruction Manual

Turbomolecular pumps
Table of Contents

Advertisement

5.4.7

ReadCounters

5
Function: Reads serial number, running time and start counter.
Transmission frame:
PC->SIM Stx
SIM->PC
PC->SIM
SIM->PC Stx
Parameter
1 to 10
11 to 20
21
22
23
Example:
Control unit serial number
Pump serial number
Pump running time
Control unit running time
Start counter
Parameter
ASCII
HEX
31 32 33 34 35 20 20 20 20 20 36 37 38 39 41 20 20 20 20 20
Parameter
ASCII
HEX
30 30 30 30 30 30 33 43 30 30 30 30 30 32 38 43 30 30 30 30 30 30 36 34
MT-94E-001-C
Page 104
STP-iS1607 Series Turbomolecular Pump
0
0
1
?
c
0
0
1
Sp
c
Item
Control unit serial number
Pump serial number
Pump running time
(Unit: minute)
Control unit running time
(Unit: minute)
Start counter
: 12345
: 6789A
: 0000003C
: 0000028C
: 00000064
1
2
3
4
5
6
7
1
2
3
4
5
21
0
0
0
0
0
0
3 C 0
Etx LRC
Ack
Parameter 1 to 23
Etx LRC
Data format
ASCII character
ASCII character
32-bits hexadecimal
coded ASCII
32-bits hexadecimal
coded ASCII
32-bits hexadecimal
coded ASCII
= 60 minutes = 1 hour
hex
= 652 minutes =10 hours and 52 minutes
hex
= 100 times
hex
8
9 10 11 12 13 14 15 16 17 18 19 20
6
7
8
9 A
22
0
0
0
0
2
Ack
Remarks
23
8 C 0
0
0
0
0
0
6
4