5.3.11 Broadcasting Command In The Rs485 Multi-Point Connection; Pump Operation Commands - Edwards STP-iS1607 Series Instruction Manual

Turbomolecular pumps
Table of Contents

Advertisement

Always assign less than 254 characters (n< 254) to the parameter so that the message is less than
256 characters.
5
Next, the preceding SIM->PC character is "Ack", the instructed query command is executed and the
SIM returns the following response (1st block).
PC->SIM
SIM->PC @ F
Then "Ack" is sent by the PC->SIM character in reaction to the response (1st block) from the SIM,
the SIM returns the following response (2nd block).
PC->SIM
SIM->PC @ F
The PC transmits "Ack" or "Nak"; then transmits the next command if necessary.

5.3.11 Broadcasting command in the RS485 multi-point connection

The START or STOP of STP pump operation command can be concurrently instructed to all the
multi-connected SIMs. Always assign 0 (HEX code "30", "30") to network frame number. Note that
there is no response from the respective SIM.
PC->SIM @ 0
SIM->PC
Parameter
1
MT-94E-001-C
Page 90
STP-iS1607 Series Turbomolecular Pump
F
Stx 0
0
1
Sp CHR C
1
2
F
Stx 0
0
2 C
1
2
254
0 Stx 0
0
1
Sp E
Item
Pump operation command
Pump operation command
START
STOP
Table 21 - Pump operation commands
C
Etb LRC
1
253
Ack or Nak
C
Etx LRC
n
Parameter 1
Etx LRC
Data Format
8-bits hexadecimal
Refer to Table 21
coded ASCII
Value
1
2
Ack or Nak
F
F
1
2
F
F
1
2
No response
Remark