5.4.8

ReadSetPoint

Function: Reads the setting value of the "Speed Set Point" and TMS temperature. The "Speed Set
Point" data is the same data as that of "ReadSpeedSetPoint".
Transmission frame:
PC->SIM Stx
0
0
SIM->PC
PC->SIM
SIM->PC Stx
0
0
Parameter
1
*1
2
TMS temperature setting
*1
Valid only with TMS specification
Example:
Speed Set Point
TMS temperature setting
Parameter
1
ASCII
0
1
F
HEX
30 31 46 34 30 30 34 36
5.4.9

ReadMotorTemp

Function: Reads the measured motor temperature.
Transmission frame:
PC->SIM Stx
0
0
SIM->PC
PC->SIM
SIM->PC Stx
0
0
STP-iS1607 Series Turbomolecular Pump
1
?
d
Etx LRC
1
Sp
d
Parameter 1 to 2
Item
Speed Set Point
(Unit: Hz)
(Unit: ºC)
: 01F4
= 500 Hz = 30,000 rpm
hex
: 0046
= 70 ºC
hex
2
4
0
0
4
6
1
?
e
Etx LRC
1
Sp
e
Parameter 1
Ack
Ack
Etx LRC
Data format
16-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
Ack
Ack
Etx LRC
5
Remarks
MT-94E-001-C