5.4.15 ReadMeasValue

Function: Reads the TMS temperature, motor temperature, motor current, measured rotational
speed, and control unit temperature. The motor temperature value is the same temperature as
"ReadMotorTemp". The measured rotational speed value is the same as "ReadMeas" parameter 2
(Measured rotational speed).
Transmission frame:
PC->SIM Stx
0
0
SIM->PC
PC->SIM
SIM->PC Stx
0
0
The HEX code of ASCII character '[' is "5B".
*1
Parameter
1
[System reservation]
*1
2
TMS temperature
3
Motor temperature
4
[System reservation]
5
6
[System reservation]
7
Measured rotational speed
8
[System reservation]
9
Control unit temperature
*1
Valid only with TMS specification
STP-iS1607 Series Turbomolecular Pump
*1
1
?
[
Etx LRC
Ack
*1
1
Sp
[
Parameter 1 to 9
Item
(Unit ºC)
(Unit ºC)
Motor current
(Unit: 0.1 A)
(Unit: Hz)
(Unit: °C)
Ack
Etx LRC
Data format
120-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
8-bits hexadecimal
coded ASCII
24-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
48-bits hexadecimal
coded ASCII
16-bits hexadecimal
coded ASCII
5
Remark
MT-94E-001-C
Page 113