Af Analyzer Frequency Accuracy To 100 Khz Performance Test 16 Record - Agilent Technologies E8285A Manual

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

AF Analyzer Frequency Accuracy to 100 kHz
Performance Test 16 Record
For test procedure, see
Performance Test 16" on page
Table 7-16
AF Analyzer Frequency Accuracy to 100 kHz Test 16 Record
21
100
1000
10000
100000

AF Analyzer Frequency Accuracy to 100 kHz Performance Test 16 Record

"AF Analyzer Frequency Accuracy to 100 kHz
Frequency
(Hz)
Lower
20.896
99.880
999.700
9997.90
99979.9
175.
Frequency Limits (Hz)
Upper
Actual
21.104
100.120
1000.300
10002.10
100020.1
Performance Test Records