CDMA Generator DUPLEX OUT Amplitude Level Accuracy Performance Test
13. Set the AWGN Power field to the values listed in the PTR for AWGN power.
See
table 7-27, "CDMA Generator DUPLEX OUT AWGN Power Level
Accuracy Test 26 Record" on page
14. Select the CDMA CALL CONTROL screen.
15. Change the Avg Power measurement field to Chan Power (Channel
Power).
16. Calibrate the channel power by selecting Calibrate in the Power Meas
field.
17. Select the CDMA GENERATOR CONTROL screen.
18. Read the channel power and record it in the PTR.
244.
Performance Tests
26
193