Performance Tests
RF Generator Level Accuracy Performance Test 5

RF Generator Level Accuracy

Performance Test 5

The purpose of this test is to verify that the Test Set meets the specification limits
in PTR (Performance Test Record)
5 Record" on page
frequencies up to 1 GHz the Test Set is set to generate levels between −10 and 125
dBm (in 5 dB steps) at it's DUPLEX OUT connector. The level is measured with
the tuned RF level function of the measuring receiver. At each frequency the
measuring receiver connection is moved to the RF IN/OUT and the level measured
from -40 to −125 dBm. As the test proceeds you may be required to recalibrate the
measuring receiver.
To extend the measurement frequency to 2 GHz the second method uses a
microwave converter and amplifier to extend the measurement range (see
6-10 on page
Setup 1
Figure 6-9
Setup 1 for Measurements to 1 GHz
Measuring Receiver
INPUT 50Ω
Procedure 1
Test Sets with firmware revision A.04.5X and above operate only at RF
NOTE
frequencies from 800 to 1000 MHz and 1700 to 2000 MHz. A text only error
message "Input value out of range" , will be displayed if invalid
frequencies are set.
Steps 1 to 5 in the following procedure apply to Setup 1 shown in
page
1. Before connecting the Test Set to the measuring receiver:
158
210. Using a measuring receiver and sensor module, at several
160).
SENSOR
158.
a. Reset the instrument.
b. Zero and calibrate the sensor module.
table 7-5, "RF Generator Level Accuracy Test
POWER
RF IN/OUT
DUPLEX OUT
DO NOT apply RF
Max power 2.5W
Max power 60 mW
when in standby
RF IN/OUT
S:\agilent\e8285\ALR\Book\chapters\perf_test.fm
figure
CALL CONTROL
UTILITIES
Answer
Print
Printer
I/O config
Help
Call/
End/
Register
Tests
Config
Previous
Message
Preset
Page
Release
Hold
Meas
CDMA SCREENS
ANALOG SCREENS
USER
reset
Cell config
Spectrum
SMS
Address
Call
Gen
Authen-
Call
Duplex
Spectrun
Local
K1'
control
control
tication
control
analyzer
K1
MS report
MS FER
Save
RX test
TX test
TX
RX test
TX test
AF
Recall
range
analyzer
K2'
K2
DATA FUNCTIONS
DATA
K3'
Ref set
Meter
Average
K3
Increment
Increment
Increment
7
8
9
Enter
10
set
x 10
Low limit
High limit
Assign
K4
4
5
6
Ghz
dBm
dB
Release
K5
1
2
3
MHz
V
%
kHz
0
.
+/-
mV
s
IBASIC
mW
reset
Yes
No
%
Hz
Shift
Cancel
On/Off
Ratio
dB
µV
W
ms
ANTENNA IN
VOLUME
AUDIO OUT
AUDIO IN
High
Low
Max 12V peak
Max
42V peak
DUPLEX OUT
Sensor Module
figure 6-9 on

Advertisement

Table of Contents
loading

Table of Contents