As some of the tests run, you may be offered the options to alter test execution
conditions by selecting:
• Loop to run the test continuously
• Pause to pause the tests
• Stp Fail (stop-failure) to stop on a failure
• Sgl Step (single-step) to pause the test after each measurement
For descriptions of the diagnostic options, refer to:
•
"Functional Diagnostics Menu" on page
•
"AF Diagnostics" on page 64
•
"RF Diagnostics" on page 65
•
"CDMA Diagnostics" on page 67
Define Test Conditions
1. On the TESTS (Main Menu) screen (see
Cond to access the TESTS (Execution Conditions) screen.
Figure 2-11
TESTS (Execution Conditions)
2. Set up the Output Results To: field. Select:
• Crt to view measurements only on the display.
• Select Printer to print the test results as well as display them on the CRT.
3. Set the Output Results For: field to All
4. Set up the If Unit-Under-Test Fails: field.
• Select Continue to continue to the next test point.
• Select Stop to pause testing at that point.
Troubleshooting
Functional Diagnostics
62.
figure
2-13), select Exec Execution