Agilent Technologies E8285A Manual page 59

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

As some of the tests run, you may be offered the options to alter test execution
conditions by selecting:
• Loop to run the test continuously
• Pause to pause the tests
• Stp Fail (stop-failure) to stop on a failure
• Sgl Step (single-step) to pause the test after each measurement
For descriptions of the diagnostic options, refer to:
"Functional Diagnostics Menu" on page
"AF Diagnostics" on page 64
"RF Diagnostics" on page 65
"CDMA Diagnostics" on page 67
Define Test Conditions
1. On the TESTS (Main Menu) screen (see
Cond to access the TESTS (Execution Conditions) screen.
Figure 2-11
TESTS (Execution Conditions)
2. Set up the Output Results To: field. Select:
• Crt to view measurements only on the display.
• Select Printer to print the test results as well as display them on the CRT.
3. Set the Output Results For: field to All
4. Set up the If Unit-Under-Test Fails: field.
• Select Continue to continue to the next test point.
• Select Stop to pause testing at that point.
Troubleshooting
Functional Diagnostics
62.
figure
2-13), select Exec Execution
59