Functional Verification
ROM-Based Diagnostics and Calibration
ROM-Based Diagnostics and Calibration
Comprehensive internal IBASIC diagnostics are provided to aid in
troubleshooting the Test Set. These diagnostic programs use external cabling
and the internal measurement and path setting capabilities of the Test Set to
diagnose itself for RF and audio hardware problems. No external test
equipment is necessary.
As the diagnostic programs execute, detected problems will be reported with a
suggested defective assembly and it's probability of failure. Refer to the
appropriate Assembly Level Repair section for more information.
1. Press the Preset key to configure instrument to default settings.
2. Press the Tests key to access to TESTS (Main Menu) screen, see
3. For Select Procedure Location: select ROM
4. For Select Procedure Filename: select SERVICE7
5. Press the K1 key. The SERVICE MENU appears.
6. Load and execute each of the following diagnostic and calibration programs:
NOTE
Disregard all other diagnostic or calibration programs that appear on the
SERVICE MENU and are not listed in step 6.
7. Load and execute PCB_CAL program:
8. Record results in the
118
on page 119
.
• Under Functional Diagnostics; select RF Modules, Analog Modulation,
and CDMA Loopback
• AF Diagnostics
• RF Diagnostics
• CDMA Diagnostics
• IQ Calibration
• IQ Demod Path Calibration
• Press the Tests key to access the TESTS (Main Menu) screen.
• For Select Procedure Location: select ROM
• For Select Procedure Filename: select PCB_CAL
• Press K1 to execute.
Table 4-1 on page
123.
S:\agilent\e8285\ALR\Book\chapters\verif.fm
Figure 4-1