Rf Generator Harmonics Spectral Purity; Performance Test 6 - Agilent Technologies E8285A Manual

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

RF Generator Harmonics Spectral Purity

Performance Test 6

The purpose of this test is to verify that the Test Set meets the specification limits
in PTR (Performance Test Record)
Purity Test 6 Record" on page
frequencies and two different levels (maximum output and minimum level vernier)
are searched for by an RF spectrum analyzer.
Setup
Figure 6-11
Spectrum Analyzer
Procedure
Test Sets with firmware revision A.04.5X and above operate only at RF
NOTE
frequencies from 800 to 1000 MHz and 1700 to 2000 MHz. A text only error
message "Input value out of range" , will be displayed if invalid
frequencies are set.
1. Set up the spectrum analyzer in accordance with its operating manual.
2. On the Test Set:
RF Generator Harmonics Spectral Purity Performance Test 6
INPUT 50Ω
a. Press
.
Preset
b. Select the CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Set the RF Offset to Off.
e. Set the Output Port field to Dupl.
f. Select the RF GENERNATOR screen.
g. Set the RF Gen Freq field to the appropriate RF frequency (based on the
firmware revision; see NOTE above) from the Performance Test Record.
h. Set the Amplitude to -10 dBm.
table 7-6, "RF Generator Harmonics Spectral
218. Harmonic signals with the carrier set to several
CALL CONTROL
Answer
Print
Call/
End/
Register
Tests
Page
Release
CDMA SCREENS
USER
Cell config
Spectrum
SMS
Call
Gen
Authen-
K1'
control
control
tication
K1
MS report
MS FER
RX test
TX test
TX
range
K2'
K2
DATA FUNCTIONS
K3'
Ref set
Meter
Average
K3
Increment
Increment
Increment
10
set
x 10
Assign
Low limit
High limit
K4
Release
K5
IBASIC
reset
Shift
Cancel
POWER
RF IN/OUT
DUPLEX OUT
ANTENNA IN
VOLUME
DO NOT apply RF
Max power 2.5W
Max power 60 mW
when in standby
DUPLEX OUT
Performance Tests
UTILITIES
Printer
I/O config
Help
Config
Previous
Message
Preset
Hold
Meas
ANALOG SCREENS
reset
Address
Call
Duplex
Spectrun
Local
control
analyzer
Save
RX test
TX test
AF
Recall
analyzer
DATA
7
8
9
Enter
Ghz
4
5
6
dBm
dB
1
2
3
MHz
V
%
kHz
0
.
+/-
mV
s
mW
Yes
No
%
Hz
On/Off
Ratio
dB
µV
W
ms
AUDIO OUT
AUDIO IN
High
Low
Max 12V peak
Max
42V peak
163

Advertisement

Table of Contents
loading

Table of Contents