Disassembly and Replaceable Parts
Before You Start
Calibration Data Download Procedure
1. Switch the Test Set's power off.
2. Remove the faulty assembly.
3. Install the replacement assembly.
4. Switch the Test Set's power on.
5. Insert the memory card.
6. Press the Tests key.
7. Set the Select Procedure Location: field to Card.
8. Set the Select Procedure Filename: field to: DNLDCAL.
9. Press the K1 key to run the test.
10.Follow the instructions on the screen.