Af Generator Ac Level Accuracy Performance Test 8 Record - Agilent Technologies E8285A Manual

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

AF Generator AC Level Accuracy
Performance Test 8 Record
For test procedure, see
on page
Table 7-8
AF Generator AC Level Accuracy Test 8 Record
AF Generator
1
1
1
1
1
1
1
1
1
1
1
1
2
2
2
2
2
2
2
2
2

AF Generator AC Level Accuracy Performance Test 8 Record

"AF Analyzer AC Level Accuracy Performance Test 12"
171.
Frequency
Level
(Hz)
(mV)
25000
4000
25000
700
25000
75
10000
4000
10000
700
10000
75
1000
4000
1000
700
1000
75
100
4000
100
700
100
75
25000
4000
25000
700
25000
75
10000
4000
10000
700
10000
75
1000
4000
1000
700
1000
75
Performance Test Records
AC Level Limits (mV)
Lower
Upper
3885.000
4115.000
682.500
717.500
70.000
80.000
3885.000
4115.000
682.500
717.500
70.000
80.000
3885.000
4115.000
682.500
717.500
70.000
80.000
3885.000
4115.000
682.500
717.500
70.000
80.000
3885.000
4115.000
682.500
717.500
70.000
80.000
3885.000
4115.000
682.500
717.500
70.000
80.000
3885.000
4115.000
682.500
717.500
70.000
80.000
Actual
223

Advertisement

Table of Contents
loading