Af Generator Frequency Accuracy Performance Test 11 Record - Agilent Technologies E8285A Manual

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Performance Test Records

AF Generator Frequency Accuracy Performance Test 11 Record

AF Generator Frequency Accuracy
Performance Test 11 Record
For test procedure, see
on page
Table 7-11
AF Generator Frequency Accuracy Test 11 Record
228
S:\agilent\e8285\ALR\Book\chapters\perf_record.fm
"AF Generator Frequency Accuracy Performance Test 11"
170.
Frequency
AF Generator
(Hz)
1
25000
1
10000
1
5000
1
2000
1
1000
1
500
1
200
1
100
1
50
1
20
2
25000
2
10000
2
5000
2
2000
2
1000
2
500
2
200
2
100
2
50
2
20
Frequency Limits (Hz)
Lower
Upper
24993.750
25006.250
9997.500
10002.500
4998.750
500.125
1999.500
2000.500
999.750
1000.250
499.875
500.125
199.950
200.050
99.975
100.025
49.988
50.012
19.995
20.005
24993.750
25006.250
9997.500
10002.500
4998.750
500.125
1999.500
2000.500
999.750
1000.250
499.875
500.125
199.950
200.050
99.975
100.025
49.988
50.012
19.995
20.005
Actual