Introduction; Equipment Required; Performance Test Record; Calibration Check Cycle - Agilent Technologies J2126A Installation And Verification Manual

Transmission test sets
Hide thumbs Also See for J2126A:
Table of Contents

Advertisement

3
Performance Tests

Introduction

64
The procedures given in this section test the performance of the
J2126/7A against the instrument specifications. The tests are
intended to be performed in the recommended order for a full
instrument calibration check. Tests can be run individually,
however it will be assumed that any preceding test would meet
specifications.

Equipment Required

Equipment required for the Performance Tests is given in
"Recommended Test Equipment" on page 65. Any equipment
which meets or exceeds the critical specification of the
equipment listed may be substituted. Recommended models are
those typically used in Agilent Service Centers. Where
appropriate, alternative models are also listed.

Performance Test Record

The results of the Performance Tests may be recorded in the
"Performance Test Record" on page 140. The Performance Test
Record lists all the tested specifications and the acceptable
limits. The results recorded at incoming inspection may be used
for comparison during periodic maintenance, troubleshooting
or after repair or adjustment.

Calibration Check Cycle

This instrument requires periodic verification of performance.
Depending on use and environmental conditions, the
instrument should be checked approximately once every 2
years, using these Performance Tests.

Instrument Option Configuration

Please refer to Chapter 1 and the instrument Specifications for
information on option configuration.
Installation and Verification Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

J2127a

Table of Contents