Functional Verification
Analog Loopback
4. Turn up the VOLUME knob to hear the 1 kHz tone.
5. SINAD meter should read >35 dB.
6. Change the Amplitude setting to –100 dBm., the SINAD meter should read
>12 dB.
7. Record results in
Table 4-1 on page
123.
SUGGESTION: Store instrument settings as a Save/Recall register (press the
Save key). Name the register SINAD.
Chapter 4