AF Analyzer AC Level Accuracy
Performance Test 12 Record
For test procedure, see
on page
Table 7-12
AF Analyzer AC Voltage Accuracy Test 12 Record
AF Analyzer AC Level Accuracy Performance Test 12 Record
"AF Analyzer AC Level Accuracy Performance Test 12"
171.
Frequency
Level
(Hz)
(mV)
15000
5000
2000
5000
200
5000
20
5000
15000
500
2000
500
200
500
20
500
15000
50
2000
50
200
50
20
50
Performance Test Records
AC Voltage Limits (mV)
Lower
Upper
4850.00
5150.00
4850.00
5150.00
4850.00
5150.00
4850.00
5150.00
485.00
515.00
485.00
515.00
485.00
515.00
485.00
515.00
48.50
51.50
48.50
51.50
48.50
51.50
48.50
51.50
Actual
229