CDMA Analyzer Average Power Level Accuracy Performance Test 28
2. On the Test Set:
a. Press
Preset
b. Select the CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Set the (Gen)-(Anl) field to 0 MHz.
e. Select CDMA CALL CONTROL.
f. Set
to 881.52 MHz.
Tune Freq
g. Set the Avg Pwr Units to mW.
3. Record the Avg Pwr reading in the PTR (see
Average Power Level Accuracy Test 28 Record" on page
4. Repeat steps 2 and 3 for each of the frequencies and levels listed in the PTR.
Performance Tests
table 7-29, "CDMA Analyzer
246).
197