Periodic Adjustments/Calibration
Setting the Timebase Latches
10. Select the Value field.
11. Rotate the knob until the counter reads as close to 10 MHz as possible.
12. Store the new DAC values (timebase calibration data) in non-volatile memory
136
S:\agilent\e8285\ALR\Book\chapters\period_adj.fm
by selecting and running the Timebase Reference Using a Counter
routine from the Periodic Calibration Menu. See
Counter" on page
132.
"Timebase Reference Using a