Functional Verification
Channel Power Calibration
Channel Power Calibration
To ensure the accuracy of the Test Set power measurement, channel power
calibration should be run before the mobile phone functional test portion of the
performance verification test plan.
1. Press the Preset key to configure instrument to default settings.
2. Connect a short RF cable between the RF IN/OUT port and DUPLEX OUT
port
3. Calibrate the channel power measurement:
• Press the (CDMA SCREENS) Call control key
• Set RF Chan Std field to US PCS
• Set Avg Power field to Chan Power
• Place the cursor on the Power Meas field and select Calibrate
• Press the knob or Enter key to begin calibration routine.
4. Record results in the
Suggestion: Store instrument settings as a Save/Recall register (press the Save
key). Name the register CPWR_CAL.
120
Table 4-1 on page
123.
S:\agilent\e8285\ALR\Book\chapters\verif.fm