Performance Tests
Oscilloscope Amplitude Accuracy Performance Test 18
4. Set the frequency as shown in the Performance Test Record (PTR). For each
178
setting, perform the following:
a. Adjust the level until the digital multimeter reads 5 V.
b. Set Controls to Main and adjust the Time/Div on the Test Set to
display 2 to 3 cycles of the waveform.
c. Set Controls to Marker and press the knob (with the cursor in the
Marker To Peak+ field) to move the marker to the peak of the
waveform.
d. Read the Lvl and compare the reading to the limits in the PTR.
S:\agilent\e8285\ALR\Book\chapters\perf_test.fm