Cdma Generator Duplex Out Amplitude Level Accuracy; Performance Test 26 - Agilent Technologies E8285A Manual

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Performance Test Records
CDMA Generator DUPLEX OUT Amplitude Level Accuracy Performance Test 26

CDMA Generator DUPLEX OUT Amplitude Level Accuracy

Performance Test 26

For test procedure, see
Accuracy Performance Test 26" on page
Table 7-26
CDMA Generator DUPLEX OUT Sector B Power Level Accuracy Test 26 Record
Table 7-27
CDMA Generator DUPLEX OUT AWGN Power Level Accuracy Test 26 Record
244
S:\agilent\e8285\ALR\Book\chapters\perf_record.fm
"CDMA Generator DUPLEX OUT Amplitude Level
RF Freq.
Level (dBm)
(MHz)
−10.50
840
−15.49
840
−11.50
1890.52
−16.49
1890.52
RF Freq.
Level (dBm)
(MHz)
−7.50
840
−9
840
−7.50
1890.52
−9
1890.52
192.
Measured Level Limits (dBm)
Lower
Upper
−11.75
−9.25
−16.74
−14.24
−12.85
−10.15
−17.84
−15.14
Measured Level Limits (dBm)
Lower
Upper
−9.25
−5.75
−10.75
−7.25
−9.35
−5.65
−10.85
−9.15
Actual
Actual