Agilent Technologies E8285A Manual page 81

Cdma mobile station test set
Hide thumbs Also See for E8285A:
Table of Contents

Advertisement

Troubleshooting
Service Screen
Module
This field is used to manually select the module that contains the circuit latches to
be selected.
Latch
This field is used to manually select the circuit latches in the module selected in the
Module field above. The latches control switch, DAC, and gain settings within the
Test Set. The value of the selected latch is displayed and changed in the Value
field. Some settings are read only.
To set a latch:
1. Use the knob to select the Module . A Choices menu will appear.
2. Move the cursor to the desired module name and press the knob to select it.
3. Use the knob to select the latch field. A Choices menu will appear.
4. Move the cursor to the desired latch name and press the knob to select it.
5. Use the knob to select the Value field. A flashing >> cursor is displayed.
6. Rotate the cursor control knob or key in a number on the keypad to modify the
value (hexadecimal).
Value (hex)
This field displays and changes the hexadecimal value for the latch shown in the
Latch field.
RAM Initialize
Selecting this field clears all SAVE registers and test programs, and any initialized
RAM disk(s), that may be in RAM. It also resets all latches to their factory
power-up configuration. If you have saved one or more instrument setups using the
SAVE function, using this function will permanently remove them.
81