CDMA Mobile Phone Functional Test
1. Connect the mobile phone to the Test Set.
2. Setup for call processing:
Knowledge of the mobile's System ID, RF Channel, and RF Channel capability
NOTE
are essential to call processing. Therefore, the mobile phone functional test
may not be successful if this information is not known.
3. Turn mobile phone on:
4. Originate Call
5. Page mobile phone:
Chapter 4
2a. Press (CDMA SCREENS) Cell config key. On the CDMA CELL SITE
CONFIGURATION screen:
• For the System ID field, if known, enter system ID number.
• For the Rgstr SID field, enter system ID number (same as System ID
field).
• Set the Pwr Up Reg field to On.
2b. Press the Config key and set the Date field for one month in advance.
2c. Press the (CDMA SCREENS) Call control key:
• Place cursor on the Avg Power field and select Chan Power
• For the RF Chan Std field enter appropriate RF channel standard
• For the Protocol field select appropriate protocol
• For the RF Channel field enter appropriate RF channel
• For the Traffic Data Mode field select Svc Opt 2
• For the Sctr A Pwr field enter –55 dBm/BW
Suggestion: Store instrument settings as a Save/Recall register (press the
Save key). Name the register CP.
• Ensure Test Set's Call Status "Transmitting" annunciator is lit.
• Mobile should obtain service with the Test Set.
• If capable, mobile will do a power-up registration within 20 seconds
indicated by the ESN appearing below the MS Database field.
• Mobile RSSI indicator shows four bars.
• Mobile may display date and time setting of Test Set.
• Dial '123' on mobile then press YES or Send
• Test Set's Connected annunciator should be lit
• Mobile should be in a connected "Svc Opt 2" state.
• Press No or End on mobile to end call.
Functional Verification
CDMA Mobile Phone Functional Test
121