Make sure the sensor module's calibration data is entered into the measuring
NOTE
receiver.
2. Connect the equipment as shown in Setup 1 whether intending to measure
3. On the measuring receiver:
4. On the Test Set:
5. For each frequency in the Performance Test Record (PTR) do the following:
frequencies to 1 GHz or 2 GHz.
a. Set the measurement mode to RF Power.
b. Set the display to log.
a. Press
.
Preset
b. Select the CONFIGURE screen.
c. Set the RF Display field to Freq.
d. Set the RF Offset to Off.
e. Set the Output Port field to Dupl.
f. Select the RF GENERNATOR screen.
g. Set the RF Gen Freq field to the appropriate RF frequency (based on the
firmware revision; see NOTE above) from the Performance Test Record.
h. Set the Amplitude to -11 dBm.
a. Set the measuring receiver to measure frequency.
b. Set the Test Set level to −15 dBm.
c. After the measuring receiver has acquired the signal, set the measuring
receiver to measure tuned RF level.
d. Measure the RF level at the levels shown in the PTR at the Test Set's
DUPLEX OUT port and compare the measured RF level to the limits. If the
measuring receiver displays the need to recalibrate, press the CALIBRATE
key and wait for calibration to be completed.
e. Move the sensor module to the Test Set's RF IN/OUT port.
f. On the Test Set set the Output Port field to RF Out and repeat the
measurements for the levels shown in the PTR and compare the measured
RF level to the limits.
g. Move the sensor module back to the Test Set's DUPLEX OUT port and set
the Output Port to Dupl.
RF Generator Level Accuracy Performance Test 5
Performance Tests
159