5. On the measuring receiver:
a. Set the frequency offset mode to enter and enable the LO frequency (27.3
Special).
b. Key in the LO frequency (in MHz) which is 1500.
6. On the Test Set, for frequencies of 1700 and 2000 MHz, measure the FM at the
deviations shown in the PTR and compare the measured deviation to the limits.
RF Generator FM Accuracy Performance Test 2
Performance Tests
151