CDMA Generator RF IN/OUT Amplitude Level Accuracy Performance Test 25
3. On the Test Set:
a. Press
.
Preset
b. Select the CONFIGURE screen and set RF display to freq.
c. Select the CDMA CALL CONTROL screen.
d. Set the RF Gen Freq field to 836.52 MHz.
e. Select the CDMA GENERATOR CONTROL screen.
f. Set the Sctr A Pwr to –20.5 dBm.
4. On the measuring receiver:
a. Set the measurement mode to RF Power.
b. Set the display to log.
c. Press the Calibrate key.
5. Set the Test Set to the frequencies and levels listed in the PTR and record the
values.
Performance Tests
191