Alarms & Troubleshooting

Flush Error Alarm (Medium Priority)

A
B
C
D
E
LED Pump Alarm Indicator Lights: Blinking Yellow
The FLUSH ERROR screen appears when the flush bag is empty or there is a
clog or a kink in the tubing between the flush bag and the pump.
Detection of this alarm condition may take up to 1 minute to occur.
Sound Intensity of Alarm: minimum 45 dBA at 1 meter
A. Refill bag as necessary.
B. Press Continue and prime the pump to prepare for next flush.
Disconnect patient from tubing before priming.
Please note: Always prime the flush tubing before the feed tubing
when using Kangaroo OMNI™ Feed Sets with Flush Bags. Priming the
flush tubing after priming the feed tubing will introduce air in the
tubing after the feeding solution, which will require the feed tubing to
be primed again to remove the air.
OR
C. If alarm continues, turn off pump and Remove the cassette from
Kangaroo OMNI™ Enteral Feeding Pump.
D. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump occlusion
sensor region.
E. Reload the cassette.