Prime Kangaroo Omni™ Enteral Feeding Pump - Cardinal Health Kangaroo Omni 385400 Instructions For Use Manual

Enteral feeding pump
Table of Contents

Advertisement

System Setup
5
5a. Auto Prime Kangaroo OMNI™ Enteral Feeding Pump Feed Only or Feed with Flush Feeding Set
To prime the Kangaroo OMNI™ Feeding Set, first fill the bag(s) with feeding solution or connect a
pre-filled container.
A
B
C
D
E
5.
Prime Kangaroo OMNI™ Enteral Feeding Pump
Note: If the Kangaroo OMNI™ Feeding Set is already primed, press the
Skip Prime button.
A. Press the Prime button.
WARNING: Patient must not be connected to Kangaroo OMNI™ Enteral
Feeding Pump when priming is performed.
Please note: When using a Feed with Flush Feeding Set, always prime
the flush tubing before the feed tubing when using Kangaroo OMNI™
Feed Sets with Flush Bags. The Auto Prime feature of Kangaroo OMNI™
will manage this for you. Priming the flush tubing after priming the feed
tubing will introduce air in the tubing after the feeding solution, which
will require the feed tubing to be primed again to remove the air.
B. Press the Auto Prime button to automatically prime the feeding set or
feed with flush set.
C. Kangaroo OMNI™ Enteral Feeding Pump will show a progress bar
indicating the automatic prime is in process. Wait for the priming
to complete.
D. Press and hold the Prime Feed button until the feeding solution just
begins to exit the distal connector.
E. Press the Done button to continue with Kangaroo OMNI™ Enteral
Feeding Pump set-up.