Sensor Error 1 Alarm (Medium Priority)

A
C. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump
valve actuator.
D. Reload the cassette.
E. If alarm continues, Replace Kangaroo OMNI™ Feeding Set.
F. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press
Power Off to stop operation and use a different Kangaroo OMNI™
Enteral Feeding Pump.
Pump Status Indicator: Blinking Yellow
The SENSOR ERROR 1 alarm appears when Kangaroo OMNI™ Enteral
Feeding Pump detects an issue with the occlusion detection sensor.
Detection of the Sensor Error 1 alarm condition pertaining to the ultrasonic
sensor may take up to 10 minutes to occur. The pump is designed to detect
degradation of sensors. If sensors degrade below acceptable levels, a
Sensor Error will result.
A. Remove the cassette.
57