Cassette Dislodged Alarm (Medium Priority)

A
B
C

Cassette Error Alarm (Medium Priority)

A
Pump Status Indicator: Blinking Yellow
The CASSETTE DISLODGED alarm appears when the cassette is not properly
loaded in Kangaroo OMNI™ Enteral Feeding Pump.
Detection of this alarm condition may take up to 1 second to occur at all
flow rates.
A. Reload cassette and press continue.
B. If alarm continues, Replace Kangaroo OMNI™ Feeding Set.
C. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press
Power Off to stop operation and use a different Kangaroo OMNI™
Enteral Feeding Pump.
Pump Status Indicator: Blinking Yellow
The CASSETTE ERROR alarm appears when Kangaroo OMNI™ Enteral
Feeding Pump has detected an unusual operating condition.
Detection of this alarm condition may take up to 31 minutes to occur
at 1 mL/hr.
Detection of this alarm condition may take up to 1 minute to occur at flow
rates greater than 50 mL/hr.s.
A. Remove the cassette and inspect it for irregularities.