Download Print this page

Cardinal Health Kangaroo ePump Quick Reference Manual

Enteral feeding pump

Advertisement

Quick Links

QUICK
REFERENCE
GUIDE
Kangaroo
ePump
TM
Enteral Feeding Pump
Below are some probable causes of alarm conditions and their corrections.
SYMPTOM PROBABLE CAUSE
Pump in Holding mode for more
Hold Error
than 10 minutes.
Occlusion in pump-patient
Flow Error
line (downstream occlusion).
Empty bag or occlusion
Feed Error
in bag-pump line
(upstream occlusion).
Pump Set
Pump Set in use beyond
Use >24 Hrs
recommended length of time.
Pump Set
Pump Set not properly
Dislodged
loaded to pump.
Battery Low
Battery charge too low.
Pump Set tubing not properly
Rotor Error
loaded on rotor, or other unusual
rotor operating condition.
General error caused by many
System Error
factors. See Manual for error list.
For additional operating instructions and information, see the operating manual for the Kangaroo
Only" Set.
© 2017 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal
Health LOGO are trademarks of Cardinal Health and may be registered in the
US and/or in other countries. 12/2017 - H9816 - WF#2098279
15 Hampshire Street
800.962.9888
Mansfield, MA 02048
508.261.8000
CORRECTIONS
Press Continue to return to Holding screen, or press and hold the
Power button to turn off. Several options are available from the
Holding screen, such as Run or Adjust Settings.
Check for and correct the occlusion in the line between pump and
patient. Replace Pump Set if error can't be resolved.
Check for empty bag and refill, or check for and correct the occlusion
in the line between the pump and feed bag. Replace Pump Set if error
can't be resolved. Check the valve pocket in the pump set loading area
for dampness or dirt.
Informational warning message blinks on upper left of screen. It's
recommended that the Pump Set be replaced.
Check black ring retainer (MISTIC) and valve; reload Pump Set and
restart pump to continue. Replace Pump Set if error can't be resolved.
Immediately plug pump into A/C outlet to recharge battery and
provide power for pump.
Check that the tubing is loaded on the rotor and that it's not
damaged, torn, etc. Reload tubing or replace Pump Set.
Power down pump and attempt restart to clear error. Call customer
service at 800.962.9888 with screen code if error can't be resolved.
Status LEDs
Red: Error
Yellow: Hold/Pause
Green: Normal Operation
Programming and Menu
Power Button
Selection Buttons
ePump Enteral Feeding Pump "Feed

Advertisement

loading

Summary of Contents for Cardinal Health Kangaroo ePump

  • Page 1 ™ Only" Set. © 2017 Cardinal Health. All Rights Reserved. CARDINAL HEALTH, the Cardinal Health LOGO are trademarks of Cardinal Health and may be registered in the US and/or in other countries. 12/2017 - H9816 - WF#2098279 15 Hampshire Street 800.962.9888...
  • Page 2 QUICK REFERENCE GUIDE EZ-MODE OPERATION Kangaroo ePump Enteral Feeding Pump “Feed Only“ Set ™ Battery Operation Unplugging the pump from the AC line will automatically put the pump on battery power. To recharge the battery, plug the pump into an AC wall outlet. Pump will automatically begin charging.