Hang Feeding Bag/Feeding Container On Iv Pole; Install The Kangaroo Omni™ Feeding Set Cassette - Cardinal Health Kangaroo Omni 385400 Instructions For Use Manual

Enteral feeding pump
Table of Contents

Advertisement

System Setup
2b. Clear previous rate/feed settings:
A
B
3
4
B
A. Press the Clear Settings button.
B. Press the Adjust Rate button or the Adjust Flush button and make
any desired adjustments. (See Set Feed/Rate section below for
additional information).
3.

Hang Feeding Bag/Feeding Container on IV Pole

CAUTION: The top of the starting volume of enteral fluids should be
10
10 inches above the top of Kangaroo OMNI™ Enteral Feeding Pump.
(25.4 cm )
A.
Hang the feeding bag or feeding container so the highest level of
enteral fluids is 10 inches (25.4 cm) above Kangaroo OMNI™ Enteral
Feeding Pump (as shown).
4.
Install the Kangaroo OMNI™ Feeding Set Cassette
CAUTION: Use only approved Kangaroo OMNI™ Feeding Sets with
Kangaroo OMNI™ Enteral Feeding Pump.
A.
Load the cassette onto Kangaroo OMNI™ Enteral Feeding Pump by
following the animated instructions on the screen. You will hear a
"click" when the cassette has been installed correctly.
A
B. The Set Loaded screen will confirm that Kangaroo OMNI™ Enteral
Feeding Pump has correctly identified the Kangaroo OMNI™ Feeding
Set. You will now be ready to prime the Kangaroo OMNI™ Feeding Set.
Optional: To change Kangaroo OMNI™ Enteral Feeding Pump
preferences, press the More Options button (see More Options section
for additional information).
B
C
D
"Click"