Alarms & Troubleshooting
Sensor Error 2 or 3 Alarm (Medium Priority)
A
B
C
D
E
Pump Status Indicator: Blinking Yellow
The SENSOR ERROR 2 or 3 alarm appears when Kangaroo OMNI™ Enteral
Feeding Pump detects an issue with its force sensor.
Detection of this alarm condition will occur immediately on power up or
cassette removal.
The Kangaroo OMNI™ pump is designed to detect degradation of sensors.
If sensors degrade below acceptable levels, a Sensor Error will result.
A. Remove the cassette.
B. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump occlusion
sensor region.
C. Reload the cassette.
D. Press Continue and restart feeding.
E. If alarm continues, Replace Kangaroo OMNI™ Feeding Set.
59