14. Do Not use Kangaroo OMNI™ Enteral Feeding Pump adjacent to or stacked with other equipment because it
could result in improper operation. If such use is necessary, Kangaroo OMNI™ Enteral Feeding Pump and the other
equipment should be observed to verify that they are operating normally.
15. Always refer to the Pediatric section on page 8 for content specific to the pediatric patient population.
16. The High Priority (Critical) audible alarm volume for the pump can be adjusted by a pump user or operator. Since
the alarm can be adjusted to a lower volume, there is a possibility that the auditory alarm signal volume can be
set to less than the ambient noise levels in the room. In this case, the noise in the room can prevent or impede
operator recognition of audible alarm signal conditions. Care should be taken to ensure that visual indicators can
be seen by the operator at all times in these cases or the medical device volume should be raised in the settings to
be above ambient noise conditions.

Cautions

1. Federal (USA) law restricts Kangaroo OMNI™ Enteral Feeding Pump for sale by or on the order of a physician.
2. Do not use Kangaroo OMNI™ Enteral Feeding Pump for delivery of any fluids or substances that are not enteral
fluids prescribed by qualified medical personnel.
3. Formula type and route of administration to be determined by the clinician. It is recommended to administer thick
formula into the stomach for optimal digestion. Enteral feeding using standard formulas and sets do not have a
restriction on anatomical placement of the access device.
4. Use of available pre-packed or prepared enteral fluids prescribed by a health care provider is recommended. Use
of homemade blenderized or liquidized foods, or non-commercially available feeding solutions may affect pump
delivery accuracy. Some enteral formulas contain chunks, seeds, or lumps that may increase the risk of occlusion.
Ensure that enteral fluids are homogenous/smooth consistency when using a feeding pump.
5. Use caution when administering thick formula in patients who cannot tolerate delivery of the prescribed volume
within 10% accuracy. This includes, but is not limited to infants, critically ill patients or those who may experience
a clinical impact due to underfeeding or overfeeding.
6. Use of thick formula feeding sets suggested for fluid levels categorized as IDDSI Level 2, 3, and 4 Drinks.
7. For use with the ENFit™ connection system. Kangaroo OMNI™ Feeding Set has the potential to misconnect with
small bore connectors of other health care applications.
8. Kangaroo OMNI™ Feeding Sets contain fluid. Accidentally disconnecting the Kangaroo OMNI™ Feeding Set from
the patient could result in a fluid leak leading to underfeeding and a potential slip hazard.
9. For optimal accuracy, the top of the starting volume of enteral fluids should be 10 inches above the top of
Kangaroo OMNI™ Enteral Feeding Pump.
10. Do not reuse feeding sets for more than one patient.
11. Should Kangaroo OMNI™ Feeding Sets require rinsing, do not manually operate the anti-free flow valve to avoid
damage. It is recommended that the Kangaroo OMNI™ Feeding Sets be rinsed using the Kangaroo OMNI™ Enteral
Feeding Pump. To rinse the feed bag, fill with rinsing fluid. Load feeding set into Kangaroo OMNI™ pump and
utilize the Kangaroo OMNI™ pump Manual Prime feature to complete rinsing of the tubing.
12. The Kangaroo OMNI™ Feeding Set should be replaced after 24 hours from initiation of feeding. This ensures that
the Kangaroo OMNI™ Enteral Feeding Pump is operating within specified parameters and prevents bacterial
growth that could be a hazard to the patient.
13. Used Kangaroo OMNI™ Feeding Sets should be disposed of in accordance with current facility procedures or local
disposal guidelines. For disposal of Kangaroo OMNI™ Enteral Feeding Pump be sure to contact local authorities
to determine the proper method of disposal of these items. Kangaroo OMNI™ Enteral Feeding Pump contains a
rechargeable lithium-ion battery.
14. Kangaroo OMNI™ Enteral Feeding Pump is designed to be used outdoors for short periods of time (no more than
24 hours). Leaving these devices outdoors for extended periods of time (exceeding 24 hours) can result in damage
and/or fading of materials.
15. Cleaning frequency and practices must be consistent with institutional policies for cleaning of non-sterile devices.
See Cleaning section for instructions on cleaning Kangaroo OMNI™ Enteral Feeding Pump. Failure to clean in
accordance with this manual could result in damage to or failure of Kangaroo OMNI™ Enteral Feeding Pump.
16. Do not store Kangaroo OMNI™ Enteral Feeding Pump, power adapter, at temperatures greater than 50ºC (122ºF).
This can damage the equipment, which will prevent Kangaroo OMNI™ Enteral Feeding Pump from operating.
6