System Setup
Intermittent Mode
Continuous Mode
B
Intermittent Mode

Turn Off Feed Complete Audio (Optional)

A feeding regime will end with either a "Feed Error" alarm or "Feed Complete" notification. The audio portion of the
"Feed Complete" notification may be disabled to avoid waking patients undergoing night time feeds.
Note: The programmed feed volume must be less than the amount of available feeding solution for the "Feed
Complete" notification to appear at the end of a feed. If the programmed feed volume is more than the available
feeding solution, Kangaroo OMNI™ Enteral Feeding Pump will alarm "Feed Error".
A
B
B. To change the rate or feed setting once feeding has started, press the
Stop button.
Note: A green screen with a moving droplet means the feeding is
in progress.
A. Press the More Options button on the Ready to Feed screen.
B. Press the down arrow button until Feed Complete Audio is highlighted.
27