anita vogel
March 16, 2025
Unable to attach the Kangaroo Joey tubing to the Jejunostomy port for enteral feeding
1 comments:
Mr. Anderson
May 14, 2025
To attach the Kangaroo Joey tubing for enteral feeding, follow these steps:
1. Ensure the tubing is pressed down and firmly seated in the loading column. 2. Grasp the black retention disk, gently wrap the tube under the rotor, and drop the disk into the slot on the right. 3. Avoid overstretching the tubing. 4. Close the blue door, making sure the finger tab on the feeding set is visible through the slot. 5. The pump display should read "set loaded" if the tubing is correctly placed. 6. Follow on-screen instructions to prime and operate the pump.
Note: The feeding port is designed to be incompatible with luer lock or IV connections to prevent misconnections.
This answer is automatically generated
james
May 20, 2025
which way to twist off a tight connection in tubing
Summary of Contents for Cardinal Health Kangaroo Joey
Page 1
Kangaroo™ Joey Enteral Feeding Pump User Guide (Feed & Hydration Mode) Pump Set Loading Procedure. 2. Confirm if you want to KEEP or 1. Locate power button on lower CLEAR any prior pump settings. right hand corner of control panel. Press to turn pump on.
Page 2
Set Priming and Pump Programming Procedure. 1. Press PRIME PUMP followed by 2. Press ADJUST FEED followed AUTO PRIME to automatically prime by FEED RATE to adjust hourly or HOLD TO PRIME to manually prime delivery rate. both feed and flush lines. Press DONE when complete.
Page 3
7. Press FLUSH VOLUME to program 8. Press buttons on left to program flush volume. Press FLUSH INTERVAL to volume of flush to be delivered. Press ENTER to confirm volume. program interval between each flush. 9. Press buttons on left to program 15 cm flush interval.
Need help?
Do you have a question about the Kangaroo Joey and is the answer not in the manual?
Questions and answers
Unable to attach the Kangaroo Joey tubing to the Jejunostomy port for enteral feeding
To attach the Kangaroo Joey tubing for enteral feeding, follow these steps:
1. Ensure the tubing is pressed down and firmly seated in the loading column.
2. Grasp the black retention disk, gently wrap the tube under the rotor, and drop the disk into the slot on the right.
3. Avoid overstretching the tubing.
4. Close the blue door, making sure the finger tab on the feeding set is visible through the slot.
5. The pump display should read "set loaded" if the tubing is correctly placed.
6. Follow on-screen instructions to prime and operate the pump.
Note: The feeding port is designed to be incompatible with luer lock or IV connections to prevent misconnections.
This answer is automatically generated
which way to twist off a tight connection in tubing