Cardinal Health Kangaroo Omni 385400 Instructions For Use Manual page 61

Enteral feeding pump
Table of Contents

Advertisement

Alarms & Troubleshooting
B
C
D
E
F
G
B. Check the tubing inside the cassette to ensure that it is free from lint or
other debris.
C. Clean and dry the Kangaroo OMNI™ Enteral Feeding Pump occlusion
sensor region. For excessive cleaning needs, avoid use of sharp objects
during the cleaning process. Use of a soft bristle brush for example may
be used.
D. Reload the cassette.
E. Press Continue and restart feeding.
F. If alarm continues, Replace Kangaroo OMNI™ Feeding Set.
G. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press
Power Off to stop operation and use a different Kangaroo OMNI™
Enteral Feeding Pump.