Alarms & Troubleshooting
Verifying Functionality of the Alarm System
A
B
C
D
E
A. Load a new Kangaroo OMNI™ Feeding Set onto Kangaroo OMNI™
Enteral Feeding Pump.
B. Leave the Kangaroo OMNI™ Feeding Set empty.
C. Run Auto Prime.
D. Upon priming, Kangaroo OMNI™ Enteral Feeding Pump will detect a
Feed Error condition and alarm.
E. Confirm the audible alarm, color display, and colored LEDs all properly
indicate a Feed Error condition.
47