Electromagnetic Conformity Declaration
Kangaroo OMNI™ Enteral Feeding Pump are intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the
user of Kangaroo OMNI™ Enteral Feeding Pump can help prevent electromagnetic interference by maintaining the minimum distance between portable and mobile
RF communications equipment (transmitters) and Kangaroo OMNI™ Enteral Feeding Pump recommended below, according to the maximum output power of the
communication equipment.
Rated maximum output
power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects, and people.
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this
device must accept any interference received, including interference that may cause undesired operation. Warning: Changes or modifications to this equipment not expressly
approved by the party responsible for compliance could void the user's authority to operate the equipment. RF exposure warning: A minimum separation distance of 20 cm
must be maintained between the device and the user or nearby persons.
This device complies with relevant Radio Standards Specifications. Operation is subject to the following two conditions: (1) This device may not cause harmful interference,
and (2) this device must accept any interference received, including interference that may cause undesired operation. Warning: Changes or modifications to this equipment
not expressly approved by the party responsible for compliance could void the user's authority to operate the equipment. RF exposure warning: A minimum separation
distance of 20 cm must be maintained between the device and the user or nearby persons.
Recommended separation distances between portable and mobile RF
communications equipment and the Kangaroo OMNI™ Enteral Feeding Pump
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
d= 1.2√P
0.12
0.38
1.2
3.8
12
m
80 MHz to 800 MHz
d=1.2√P
0.12
0.38
1.2
3.8
12

FCC STATEMENT

IC STATEMENT

800 MHz to 2,5 GHz
d = 2.3√P
0.23
0.73
2.3
7.3
23
74