Manual Certification

The following procedure can be used to manually certify Kangaroo OMNI™ Enteral Feeding Pump. It is
recommended to conduct a performance test every 2 years, or as recommended by facility protocol. Load a standard
Kangaroo OMNI™ Feeding Set filled with at least 50 mL of water. The Kangaroo OMNI™ Feeding Set should be
pre-primed, with water visible in the downstream tubing.
A
B. Use the arrow buttons to highlight Timed Pause.
C. Press the Select button to toggle between On or Off.
D. Press the Done button.
E. Press Restart Pump button to exit Biotech mode and return to normal
operating mode.
A. Press Test Pump button.