C

Adjust Brightness:

This feature adjusts the brightness of the display backlight.
A
B
C

Interruption Monitor Screen

C. Press the OK button to return to More Options.
A. In the More Options screen, use the arrow keys to select Adjust
Brightness, and press the Select button.
B. Press the + and – buttons to input the desired brightness.
C. Press the OK button to return to More Options.
Note: This screen will only appear if Interruption Monitor is enabled in
the Biotech menu.
If a continuous feeding regimen has been configured, Kangaroo OMNI™
Enteral Feeding Pump can notify users of its non-feeding time. This
feature is intended to assist practitioners of volume-based feeding.
This NOTICE screen will be displayed if the pump has been turned off,
stopped or flushing for longer than 30 minutes.
The Interrupted Time and Feed Missed totals will continue to
accumulate until cleared by the user or the programmed feed rate
is changed.
Note: A new message will appear after every additional 30 minutes of
interrupted feed time.