Alarms & Troubleshooting
F
G

Rotor Stuck Alarm (Medium Priority)

A
B
F. If alarm continues, replace Kangaroo OMNI™ Feeding Set.
G. If alarm continues after replacing Kangaroo OMNI™ Feeding Set, press
Power Off to stop operation and use a different Kangaroo OMNI™
Enteral Feeding Pump.
LED Pump Alarm Indicator Lights: Blinking Yellow
The ROTOR STUCK alarm appears when Kangaroo OMNI™ Enteral Feeding
Pump detects an unusual amount of resistance on the rotor, typically
caused by enteral fluids or contaminant build-up.
Detection of this alarm condition may take up to 2 minutes to occur.
Sound Intensity of Alarm: minimum 45 dBA at 1 meter
WARNING: Do not clean while plugged into an electrical outlet.
A. Clean the rotor by wiping it down with a wet cloth.
B. If alarm continues, press Power Off to stop operation and use a different
Kangaroo OMNI™ Enteral Feeding Pump.
52