Performing OPEN Compensation

OPEN Compensation corrects for stray admittance due to the test xture. How to perform an
OPEN compensation sequence for the 16192A is described as an example. The procedures for
the 16191A and 16193A are shown in Appendix C.
1. Set the DUT on the test xture just as you set the
shorting device in the SHORT compensation.

OPEN Compensation Key Sequence

When the test xture is ready for the OPEN compensation sequence, perform the following
procedure:
1. Press
NNNNNNNNNNNNNN
OPEN .
After the OPEN compensation sequence is done, the
2. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
DONE: COMPEN .
3. Verify that \ Cmp " appears on the left of the display.
Note
You can perform the desired xture compensation individually. To do this,
press
desired compensation.