Limit Test

You can perform limit test for only primary or secondary parameters.
1. Press
, and then
4
Test Setup
5
2. Press
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
PRI LIMIT on OFF (primary parameters) or
parameters) to toggle it
When either of the limit test keys is ON, measured data outside the limits are output and
displayed as follows:
On-screen FAIL Message
Beeper (ON/OFF selectable)
GPIB Command :DATA? LFA , :DATA? LLIS , :DATA? LMAR
Note
The frequency (stimulus value) of limit table for setting limits can be set
independently of the frequency set in the list table.
If the frequency of the limit table does not equal to the measurement
frequency of list table, limit tests for all measurements can be performed using
interpolation of the limit values.
For example, if only one point is de ned in limit table, limit test is performed
with the constant upper limit and lower limit for all measurements.
4-6
4286A with Chip Handler
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LIMIT TEST MENU .
NNNNNNNNNNNNNNNNNNNN
ON off .
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SEC LIMIT on OFF (secondary