Table 4-12. Signal De nition (Mode 2) (continued)
Pin
Signal
No.
30
/INDEX
31
/EOM
32
See Table 4-9.
to
36
Input/
Output
Output
Analog measurement end signal. Signal de nition varies depending on the
sweep mode used, as shown below.
On Sweep Mode
Is output at the end of analog measurement of the last sweep point. You
can connect the handler to a new DUT when this signal is received by the
handler.
On Point Mode
Is output at the end of analog measurement of each sweep point. You can
start measurement of the next sweep point when this signal is received
by the handler.
Note that measured data will be available only when /EOM is output,
regardless of the sweep mode used.
Output
Measurement cycle end signal. Signal de nition varies depending on the
sweep mode used, as shown below.
On Sweep Mode
Is output at the end of sweep measurement. Measured data and
comparator signals are valid when this signal is output.
On Point Mode
Is output when valid data are obtained after measurement (including time
required for the comparator to make comparisons) of each sweep point.
Comparator signals are valid when /EOM is output at the end of
measurement of the last sweep point.
|
Same as with mode 1. (See Table 4-9.)
De nition and Description
4286A with Chip Handler
4-31