15. Press
LOAD .
After a LOAD calibration sequence of about 20 seconds, the
underlined.
16. Disconnect the 50 termination.
17. Connect the low-loss capacitor to the APC-7 R
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
18. Press
LOW-LOSS CAPACITOR .
After a low-loss capacitor calibration sequence of about 20 seconds, the
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
LOW-LOSS CAPACITOR softkey label is underlined.
19. Disconnect low-loss capacitor.
20. Press
NNNNNNNNNNNNNNNNNNNNNNNNNN
DONE:CAL .
21. Verify that \ CO+ " (when you have selected
NNNNNNNNNNNNNN
(
USER ) is displayed on the left of the screen.
The calibration data is erased when the 4286A is turned o .
3-10
Basic Measurement Procedures
Figure 3-11.
Figure 3-12.
NNNNNNNNNNNNNN
LOAD softkey label is
connector as shown in Figure 3-12.
NNNNNNNNNNNNNNNNN
FIXED ) or \ Co+ " (when you have selected