See Function Reference for detailed description of measurement time.
Comparison and output require about 0.5ms
The display time is about 3ms+6ms/points.
Pin Assignment and Signal De nitions (Mode 2)
The relationship between the primary/secondary parameter values and signals /F1REJ to
/F10REJ and /DBIN0 to /DBIN3 is as follows:
Time
T1
Trigger pulse width
T2
Measurement start delay
T3
Wait time from /EOM output to
triggering
T3 represents the
minimum duration of time
4286A to function properly. Min. and Max. represents the time
required respectively when the most simple and most complex
measurement conditions are set. Note that when you specify
excessively short time, /EXT TRIG may be ignored, causing the
4286A to malfunction.
The items which can a ect on T3 are listed below. T3 gets
the items are ON and/or the number of frequency points increases.
a. Number of measurement points
b. Setting the beeper ON/OFF
c. Setting limit test ON/OFF, and number of frequency points
d. Setting BIN sorting ON/OFF, and number of frequency points
The beeper signi cantly a ects T3, adding 100ms when it is ON.
Min.

1
s
|
5ms
1
required for the
Max.
|

600
s
600ms
if
longer
4286A with Chip Handler
4-27