Type 8619
Technical data
6.10.6
Specifications of the conductivity module "COND"
Resistance measurement (without
conductivity sensor connected)
Power consumption
Type of conductivity probe
Conductivity measurement (with
connected conductivity sensor)
• Measurement range
• Measurement resolution
• Systematic variation in the measurement
Resistivity measurement
(with connected conductivity sensor)
• Measurement range
• Measurement resolution
• Systematic variation in the measurement
(without sensor)
Temperature measurement
• Measurement range
• Measurement resolution
• Systematic variation in the measurement
• Temperature sensor type
6.10.7
Specifications of the Ethernet module M1
Power consumption
Supported network protocols
LEDs
Electrical connection
5 Ω...1 MΩ
0.25 VA
With 2 or 4 electrodes; the specifications of Bürkert cells are
described in the related operating instructions.
• 0.000 µS/cm...2 S/cm (depends on the conductivity sensor)
• 10–9 S/cm
• ±0.5 % of the measured value + conductivity sensor error
• 0.500 Ω.cm...100 MΩ.cm (depends on the conductivity sensor)
• 10–1 Ω.cm
• ±0.5 % of the measured value + conductivity sensor error
• –40 °C...+200 °C
• ±0.1 °C
• ±1 °C + temperature probe error
• Pt100 or Pt1000, with 2 or 3 wires
2.2 VA
• Modbus TCP
• PROFINET
• EtherNet/IP
• 2 Link/Act LED (yellow)
• 2 Link LEDs (green)
• 2 ports for an RJ45 connector
English
35